UA Strip Analyzer Errors

UA Unavailable/Unknown (the strip analyzer will always show "offline")

  • See if the UA screen is frozen

  • If frozen, reset it with a paper clip by inserting into the small hole in the back and letting it turn off completely

  • Press the grey button to turn it back on

  • Go to Settings on the strip analyzer machine and make sure USB is on

  • Close the Fuse website and restart it

  • If the UA strip analyzer is still showing unknown, go to "Tools" on the SA screen

  • Click "Options", "Network" - make sure adapter is upper NW port and DHCP is clicked

  • Go to "Host", click disconnect and then connect, "Apply" and hit "ok"

  • Go to the VetScan UA Strip Analyzer machine

  • UA Settings, disconnect then connect

  • Apply, hit ok

  • Close FUSE website and reopen

  • UA analyzer should say "Offline"

For direct help via email, ashley.ritchie@zoetis.com