UA Strip Analyzer Errors
UA Unavailable/Unknown (the strip analyzer will always show "offline")
See if the UA screen is frozen
If frozen, reset it with a paper clip by inserting into the small hole in the back and letting it turn off completely
Press the grey button to turn it back on
Go to Settings on the strip analyzer machine and make sure USB is on
Close the Fuse website and restart it
If the UA strip analyzer is still showing unknown, go to "Tools" on the SA screen
Click "Options", "Network" - make sure adapter is upper NW port and DHCP is clicked
Go to "Host", click disconnect and then connect, "Apply" and hit "ok"
Go to the VetScan UA Strip Analyzer machine
UA Settings, disconnect then connect
Apply, hit ok
Close FUSE website and reopen
UA analyzer should say "Offline"
For direct help via email, ashley.ritchie@zoetis.com